The determination of the spring constants of atomic force microscope (AFM) cantilevers is of fundamental importance to users of the AFM. In this paper, a fast and nondestructive method for the evaluation of the spring constant which relies solely on the determination of the unloaded resonant frequency of the cantilever, a knowledge of its density or mass, and its dimensions is proposed. This is in contrast to the method of Cleveland et al. [Rev. Sci. Instrum. 64, 403 (1993)], which requires the attachment of masses to the cantilever in the determination of the spring constant. A number of factors which can influence the resonant frequency are examined, in particular (i) gold coating, which can result in a dramatic variation in the resonant frequency, for which a theoretical account is presented and (ii) air damping which, it is found, leads to a shift of ∼4% in the resonant frequency down on its value in a vacuum. Furthermore, the point of load on the cantilever is found to be extremely important, since a small variation in the load point can lead to a dramatic variation in the spring constant. Theoretical results that account for this variation, which, it is believed will be of great practical value to the users of the AFM, are given.

Force Microscopy Techniques and ApplicationsMechanical and Optical ResonatorsNear-Field Optical MicroscopyCantileverSpring (device)Constant (computer programming)CalibrationAtomic force microscopyMaterials scienceCoatingNon-contact atomic force microscopyMicroscopePoint (geometry)
Citations
973
FWCI
11.83
field-weighted impact
References
27
Percentile
99%
vs. same field & year
Citations per year
Cited by
Micromechanical cantilever-based biosensors
Sensors and Actuators B Chemical · 2001 · 728 citations
Short cantilevers for atomic force microscopy
Review of Scientific Instruments · 1996 · 490 citations
Calibration of rectangular atomic force microscope cantilevers
Review of Scientific Instruments · 1999 · 2,097 citations
Normal and torsional spring constants of atomic force microscope cantilevers
Review of Scientific Instruments · 2004 · 538 citations
Calibration of frictional forces in atomic force microscopy
Review of Scientific Instruments · 1996 · 596 citations
Cantilever-based biosensors
Analytical and Bioanalytical Chemistry · 2004 · 409 citations
References
Vibration problems in engineering
Bulletin of Miscellaneous Information (Royal Gardens Kew) · 1928 · 3,456 citations
Calibration of atomic-force microscope tips
Review of Scientific Instruments · 1993 · 4,148 citations
Formulas for Stress and Strain
Journal of Applied Mechanics · 1976 · 1,928 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.