Scinovex
articleTop 10% cited

A robot with improved absolute positioning accuracy for CT guided stereotactic brain surgery

IEEE Transactions on Biomedical Engineering · 1988 · Vol. 35(2) · pp. 153–160
Y. S. KwohJung-Fu HouEdmond JonckheereS. Hayati

Abstract

The use of a Unimation Puma 200 robot, properly interfaced with a computerized tomographic (CT) scanner and with a probe guide mounted at its end effector for CT-guided brain tumor biopsis is discussed. Once the target is identified on the CT picture, a simple command allows the robot to move to a position such that the end-effector probe guide points toward the target. This results in a procedure faster than one using a manually adjustable frame. Probably the most important advantage, as is shown, is the improved accuracy that can be achieved by proper calibration of the robot.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

Soft Robotics and ApplicationsRobotic Mechanisms and DynamicsRobotics and Sensor-Based LocalizationComputer visionScannerArtificial intelligenceRobotCalibrationRobot end effectorComputer sciencePosition (finance)Frame (networking)Mathematics

MeSH terms

BrainBrain MappingHumansRoboticsStereotaxic TechniquesTomography, X-Ray Computed
Citations
1,243
FWCI
9.71
field-weighted impact
References
5
Percentile
99%
vs. same field & year
Citations per year
Cited by
30 Years of Robotic Surgery
World Journal of Surgery · 2016 · 447 citations
Robotic Surgery
Annals of Surgery · 2003 · 1,236 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.