Abstract

The main advantages of proton therapy are the reduced total energy deposited in the patient as compared to photon techniques and the finite range of the proton beam. The latter adds an additional degree of freedom to treatment planning. The range in tissue is associated with considerable uncertainties caused by imaging, patient setup, beam delivery and dose calculation. Reducing the uncertainties would allow a reduction of the treatment volume and thus allow a better utilization of the advantages of protons. This paper summarizes the role of Monte Carlo simulations when aiming at a reduction of range uncertainties in proton therapy. Differences in dose calculation when comparing Monte Carlo with analytical algorithms are analyzed as well as range uncertainties due to material constants and CT conversion. Range uncertainties due to biological effects and the role of Monte Carlo for in vivo range verification are discussed. Furthermore, the current range uncertainty recipes used at several proton therapy facilities are revisited. We conclude that a significant impact of Monte Carlo dose calculation can be expected in complex geometries where local range uncertainties due to multiple Coulomb scattering will reduce the accuracy of analytical algorithms. In these cases Monte Carlo techniques might reduce the range uncertainty by several mm.

Radiation Therapy and DosimetryAdvanced Radiotherapy TechniquesRadiation Detection and Scintillator TechnologiesMonte Carlo methodProton therapyRange (aeronautics)ProtonComputational physicsPhysicsStatistical physicsMaterials scienceNuclear physicsMathematics

MeSH terms

HumansMonte Carlo MethodProtonsRadiotherapyScattering, RadiationTomography, X-Ray ComputedUncertaintyProton Therapy

Funding

  • National Institutes of Health
  • University of Texas MD Anderson Cancer Center
  • National Cancer Institute
Citations
1,266
FWCI
45.66
field-weighted impact
References
83
Percentile
100%
vs. same field & year
Citations per year
References
The calibration of CT Hounsfield units for radiotherapy treatment planning
Physics in Medicine and Biology · 1996 · 1,062 citations
Relative biological effectiveness (RBE) values for proton beam therapy
International Journal of Radiation Oncology*Biology*Physics · 2002 · 1,327 citations
Radiological Use of Fast Protons
Radiology · 1946 · 1,350 citations
Geant4—a simulation toolkit
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2003 · 23,965 citations
Related articles
Range uncertainties in proton therapy and the role of Monte Carlo simulations
Physics in Medicine and Biology · 2012 · 1,266 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Range uncertainties in proton therapy and the role of Monte Carlo simulations · Scinovex