Scinovex
reviewTop 1% cited

<i>In vivo</i>proton range verification: a review

Physics in Medicine and Biology · 2013 · Vol. 58(15) · pp. R131–R160
Antje KnopfAntony Lomax

Abstract

Protons are an interesting modality for radiotherapy because of their well defined range and favourable depth dose characteristics. On the other hand, these same characteristics lead to added uncertainties in their delivery. This is particularly the case at the distal end of proton dose distributions, where the dose gradient can be extremely steep. In practice however, this gradient is rarely used to spare critical normal tissues due to such worries about its exact position in the patient. Reasons for this uncertainty are inaccuracies and non-uniqueness of the calibration from CT Hounsfield units to proton stopping powers, imaging artefacts (e.g. due to metal implants) and anatomical changes of the patient during treatment. In order to improve the precision of proton therapy therefore, it would be extremely desirable to verify proton range in vivo, either prior to, during, or after therapy. In this review, we describe and compare state-of-the art in vivo proton range verification methods currently being proposed, developed or clinically implemented.

Radiation Therapy and DosimetryAdvanced Radiotherapy TechniquesRadiation Detection and Scintillator TechnologiesProton therapyRange (aeronautics)Hounsfield scaleProtonIn vivoNuclear medicineMedical physicsComputer scienceMaterials scienceBiomedical engineering

MeSH terms

Diagnostic ImagingHumansProtonsRadiotherapy Planning, Computer-AssistedUncertaintyPhysical PhenomenaProton Therapy
Citations
520
FWCI
27.96
field-weighted impact
References
179
Percentile
100%
vs. same field & year
Citations per year
References
Range uncertainties in proton therapy and the role of Monte Carlo simulations
Physics in Medicine and Biology · 2012 · 1,266 citations
The calibration of CT Hounsfield units for radiotherapy treatment planning
Physics in Medicine and Biology · 1996 · 1,062 citations
Relative biological effectiveness (RBE) values for proton beam therapy
International Journal of Radiation Oncology*Biology*Physics · 2002 · 1,327 citations
Radiological Use of Fast Protons
Radiology · 1946 · 1,350 citations
Intensity modulation methods for proton radiotherapy
Physics in Medicine and Biology · 1999 · 533 citations
The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy
International Journal of Radiation Oncology*Biology*Physics · 2000 · 1,511 citations
Charged hadron tumour therapy monitoring by means of PET
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2004 · 384 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.