Abstract

Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p < 0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.

Orthopaedic implants and arthroplastyAdvanced materials and compositesOrthopedic Infections and TreatmentsHip resurfacingTotal hip replacementHip replacementMaterials scienceTotal hip arthroplastyMetallurgyMedicineArthroplastySurgery

MeSH terms

AdultAgedAged, 80 and overFemaleHip JointHumansMaleMetalsMiddle AgedPain, PostoperativePostoperative PainProsthesis DesignProsthesis FailureReoperationTime Factors
Citations
731
FWCI
74.74
field-weighted impact
References
41
Percentile
100%
vs. same field & year
Citations per year
Cited by
References
Assessing activity in joint replacement patients
The Journal of Arthroplasty · 1998 · 895 citations
Dislocation After Total Hip Arthroplasty Causes and Prevention
Clinical Orthopaedics and Related Research · 1990 · 675 citations
Pseudotumours associated with metal-on-metal hip resurfacings
Journal of Bone and Joint Surgery - British Volume · 2008 · 1,035 citations
Metal-on-Metal Bearings and Hypersensitivity in Patients with Artificial Hip Joints
Journal of Bone and Joint Surgery · 2005 · 1,126 citations
Reducing the risk of dislocation after total hip arthroplasty
Journal of Bone and Joint Surgery - British Volume · 2005 · 561 citations
Traumatic Arthritis of the Hip after Dislocation and Acetabular Fractures
Journal of Bone and Joint Surgery · 1969 · 5,607 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.