Scinovex
articleTop 10% cited

Time of flight in PET revisited

IEEE Transactions on Nuclear Science · 2003 · Vol. 50(5) · pp. 1325–1330
W.W. Moses

Abstract

PET scanners based on LSO have the potential for significantly better coincidence timing resolution than the 6 ns FWHM typically achieved with BGO. This study analyzes the performance enhancements made possible by improved timing as a function of the coincidence time resolution. If 500 ps FWHM coincidence timing resolution can be achieved in a complete PET camera, the following four benefits can be realized for whole-body FDG imaging: 1) the random event rate can be reduced by using a narrower coincidence timing window, increasing the peak NECR by /spl sim/50%; 2) using time-of-flight (TOF) in the reconstruction algorithm will reduce the noise variance by a factor of 5; 3) emission and transmission data can be acquired simultaneously, reducing the total scan time; and 4) axial blurring can be reduced by using TOF to determine the correct axial plane of origin for each event. While TOF was extensively studied in the 1980s, practical factors limited its effectiveness at that time and little attention has been paid to timing in PET since then. As these potential improvements are substantial and the advent of LSO PET cameras gives us the means to obtain them without other sacrifices, efforts to improve PET timing should resume after their long dormancy.

Medical Imaging Techniques and ApplicationsRadiation Detection and Scintillator TechnologiesAtomic and Subatomic Physics ResearchCoincidenceTime of flightPhysicsFull width at half maximumOpticsPet imagingNoise (video)Image resolutionDead timePositron emission tomography
Citations
333
FWCI
11.26
field-weighted impact
References
35
Percentile
99%
vs. same field & year
Citations per year
References
A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography
IEEE Transactions on Nuclear Science · 1986 · 461 citations
Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillator
IEEE Transactions on Nuclear Science · 1992 · 835 citations
Exact and approximate rebinning algorithms for 3-D PET data
IEEE Transactions on Medical Imaging · 1997 · 749 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Time of flight in PET revisited · Scinovex