Scinovex
article Open AccessTop 1% cited

The Relationship Between Stress Shielding and Bone Resorption Around Total Hip Stems and the Effects of Flexible Materials

Clinical Orthopaedics and Related Research · 1992 · Vol. 274(274) · pp. 124???134–124???134
R. HuiskesHarrie WeinansBert van Rietbergen

Abstract

Bone resorption around hip stems is a disturbing phenomenon, although its clinical significance and its eventual effects on replacement longevity are as yet uncertain. The relationship between implant flexibility and the extent of bone loss, frequently established in clinical patient series and animal experiments, does suggest that the changes in bone morphology are an effect of stress shielding and a subsequent adaptive remodeling process. This relationship was investigated using strain-adaptive bone-remodeling theory in combination with finite element models to simulate the bone remodeling process. The effects of stem material flexibility, bone flexibility, and bone reactivity on the process and its eventual outcome were studied. Stem flexibility was also related to proximal implant/bone interface stresses. The results sustain the hypothesis that the resorptive processes are an effect of bone adaptation to stress shielding. The effects of stem flexibility are confirmed by the simulation analysis. It was also established that individual differences in bone reactivity and mechanical bone quality (density and stiffness) may account for the individual variations found in patients and animal experiments. Flexible stems reduce stress shielding and bone resorption. However, they increase proximal interface stresses. Hence, the cure against bone resorption they represent may develop into increased loosening rates because of interface debonding and micromotion. The methods presented in this paper can be used to establish optimal stem-design characteristics or check the adequacy of designs in preclinical testing procedures.

Orthopaedic implants and arthroplastyOrthopedic Infections and TreatmentsTotal Knee Arthroplasty OutcomesStress shieldingFlexibility (engineering)Bone resorptionMedicineBone remodelingResorptionImplantStress (linguistics)Biomedical engineeringDentistry

MeSH terms

Biomechanical PhenomenaBone ResorptionFemurHip ProsthesisHumansModels, BiologicalProsthesis DesignStress, MechanicalTime FactorsBone DensityBone Remodeling
Citations
1,204
FWCI
13.52
field-weighted impact
References
20
Percentile
99%
vs. same field & year
Citations per year
References
Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results
Journal of Bone and Joint Surgery - British Volume · 1987 · 1,760 citations
Adaptive bone-remodeling theory applied to prosthetic-design analysis
Journal of Biomechanics · 1987 · 1,149 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.