Scinovex
articleTop 10% cited

Beam hardening in X-ray reconstructive tomography

Physics in Medicine and Biology · 1976 · Vol. 21(3) · pp. 390–398
Rodney A. BrooksGiovanni Di Chiro

Abstract

As a polychromatic X-ray beam passes through matter, low energy photons are preferentially absorbed, and the (logarithmic) attenuation is no longer a linear function of absorber thickness. This leads to various artifacts in reconstructive tomography. If a water bag is used, the nonlinear attenuation in bone causes a distortion of the bone values and a spill-over inside the skull, or 'pseudo-cortex' artifact. If no water bag is used, there is an additional effect due to the varying thickness of soft tissue which causes a depression of interior values, or 'cupping'. Both artifacts can be remedied by additional prefiltering of the beam and by applying a linearization correction to the detector outputs.

Medical Imaging Techniques and ApplicationsAdvanced X-ray and CT ImagingAdvanced X-ray Imaging TechniquesAttenuationTomographyBeam (structure)OpticsLinearizationImaging phantomMaterials scienceArtifact (error)PhysicsWedge (geometry)

MeSH terms

Diagnosis, Computer-AssistedHistory of MedicineTomography, X-Ray
Citations
588
FWCI
6.48
field-weighted impact
References
11
Percentile
97%
vs. same field & year
Citations per year
References
The Fourier reconstruction of a head section
IEEE Transactions on Nuclear Science · 1974 · 2,155 citations
The Fourier Transform and Its Applications
American Journal of Physics · 1966 · 5,454 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.