Scinovex
articleTop 1% cited

Label-free protein assay based on a nanomechanical cantilever array

Nanotechnology · 2002 · Vol. 14(1) · pp. 86–90
Youri ArntzJoachim SeeligH.P. LangJ. ZhangPatrick HunzikerJ.-P. RamseyerErnst MeyerMartin HegnerCh. Gerber

Abstract

We demonstrate continuous label-free detection of two cardiac biomarker proteins (creatin kinase and myoglobin) using an array of microfabricated cantilevers functionalized with covalently anchored anti-creatin kinase and anti-myoglobin antibodies. This method allows biomarker proteins to be detected via measurement of surface stress generated by antigen–antibody molecular recognition. Reference cantilevers are used to eliminate thermal drifts, undesired chemical reactions and turbulences from injections of liquids by calculating differential deflection signals with respect to sensor cantilevers. The sensitivity achieved for myoglobin detection is below 20 µg ml−1. Both myoglobin and creatin kinase could be detected independently using cantilevers functionalized with the corresponding antibodies, in unspecific protein background. This approach permits the use of up to seven different antigen–antibody reactions simultaneously, including an additional thermomechanical and chemical in situ reference. Applications lie in the field of early and rapid diagnosis of acute myocardial infarction.

Mechanical and Optical ResonatorsForce Microscopy Techniques and ApplicationsAdvanced MEMS and NEMS TechnologiesMyoglobinCantileverMaterials scienceBiosensorBiomarkerCovalent bondNanotechnologyBiophysicsChemistryBiochemistry

Funding

  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Citations
408
FWCI
13.18
field-weighted impact
References
18
Percentile
99%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.