A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals

IEEE Transactions on Nuclear Science · 1999 · Vol. 46(3) · pp. 468–473
A.P. JeavonsRebecca ChandlerC.A.R. Dettmar

Abstract

A HIDAC-PET camera consisting essentially of 5 million 0.5 mm gas avalanching detectors has been constructed for small-animal imaging. The particular HIDAC advantage-a high 3D spatial resolution-has been improved to 0.95 mm fwhm and to 0.7 mm fwhm when reconstructing with 3D-OSEM methods incorporating resolution recovery. A depth-of-interaction resolution of 2.5 mm is implicit, due to the laminar construction. Scatter-corrected sensitivity, at 8.9 cps/kBq (i.e. 0.9%) from a central point source, or 7.2cps/kBq (543cps/kBq/cm/sup 3/) from a distributed (40 mm diameter, 60 mm long) source is now much higher than previous, and other, work. A field-of-view of 100 mm (adjustable to 200 mm) diameter by 210 mm axially permits whole-body imaging of small animals, containing typically 4 MBqs of activity, at 40 kcps of which 16% are random coincidences, with a typical scatter fraction of 44%. Throughout the field-of-view there are no positional distortions and relative quantitation is uniform to /spl plusmn/3.5%, but some variation of spatial resolution is found. The performance demonstrates that HIDAC technology is quite appropriate for small-animal PET cameras.

Medical Imaging Techniques and ApplicationsRadiation Detection and Scintillator TechnologiesAdvanced Fluorescence Microscopy TechniquesImage resolutionFull width at half maximumField of viewResolution (logic)Materials sciencePhysicsDetectorOpticsNuclear medicineComputer science

Funding

  • Medical Research Council
Citations
257
FWCI
15.29
field-weighted impact
References
23
Percentile
99%
vs. same field & year
Citations per year
References
MicroPET: a high resolution PET scanner for imaging small animals
IEEE Transactions on Nuclear Science · 1997 · 615 citations
Initial results from the Sherbrooke avalanche photodiode positron tomograph
IEEE Transactions on Nuclear Science · 1996 · 261 citations
Accelerated image reconstruction using ordered subsets of projection data
IEEE Transactions on Medical Imaging · 1994 · 3,675 citations
Analytic 3D image reconstruction using all detected events
IEEE Transactions on Nuclear Science · 1989 · 761 citations