MicroPET: a high resolution PET scanner for imaging small animals

IEEE Transactions on Nuclear Science · 1997 · Vol. 44(3) · pp. 1161–1166
Simon R. CherryYiping ShaoRobert W. SilvermanK. MeadorsStefan SiegelArion F. ChatziioannouJohn W. YoungW. J. JonesJ.C. MoyersD.F. NewportA. BoutefnouchetT.H. FarquharM. AndreacoMichael J. PaulusD.M. BinkleyR. NuttMichael E. Phelps

Abstract

MicroPET is a high resolution positron emission tomography (PET) scanner designed for imaging small laboratory animals. It consists of a ring of 30 position-sensitive scintillation detectors, each with an 8/spl times/8 array of small lutetium oxyorthosilicate (LSO) crystals coupled via optical fibers to a multi-channel photomultiplier tube. The detectors have an intrinsic resolution averaging 1.68 mm, an energy resolution between 15 and 25% and 2.4 ns timing resolution at 511 keV. The detector ring diameter of microPET is 17.2 cm with an imaging field of view of 112 mm transaxially by 18 mm axially. The scanner has no septa and operates exclusively in 3D mode. Reconstructed image resolution 1 cm from the center of the scanner is 2.0 mm and virtually isotropic, yielding a volume resolution of 8 mm/sup 3/. For comparison, the volume resolution of state-of-the-art clinical PET systems is in the range of 50-75 mm/sup 3/. Initial images of phantoms have been acquired and are reported. A computer controlled bed is under construction and will incorporate a small wobble motion to improve spatial sampling. This is projected to further enhance spatial resolution. MicroPET is the first PET scanner to incorporate the new scintillator LSO and to our knowledge is the highest resolution multi-ring PET scanner currently in existence.

Medical Imaging Techniques and ApplicationsRadiation Detection and Scintillator TechnologiesAdvanced Radiotherapy TechniquesScannerImage resolutionOpticsResolution (logic)PhysicsPhotomultiplierScintillatorDetectorField of viewPositron emission tomography
Citations
615
FWCI
24.78
field-weighted impact
References
29
Percentile
100%
vs. same field & year
Citations per year
Cited by
A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals
IEEE Transactions on Nuclear Science · 1999 · 257 citations
Molecular Imaging
Radiology · 2001 · 1,290 citations
References
A Multicrystal Two Dimensional BGO Detector System for Positron Emission Tomography
IEEE Transactions on Nuclear Science · 1986 · 461 citations
Initial results from the Sherbrooke avalanche photodiode positron tomograph
IEEE Transactions on Nuclear Science · 1996 · 261 citations
Simple charge division readouts for imaging scintillator arrays using a multi-channel PMT
IEEE Transactions on Nuclear Science · 1996 · 311 citations
Analytic 3D image reconstruction using all detected events
IEEE Transactions on Nuclear Science · 1989 · 761 citations
3-D phantom to simulate cerebral blood flow and metabolic images for PET
IEEE Transactions on Nuclear Science · 1990 · 354 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.