article Open Access
Catastrophic failure of a pyrocarbon hemiarthroplasty from glenoid metal anchors
International Journal of Case Reports in Orthopaedics · 2024 · Vol. 6(1) · pp. 135–139
Abstract
Shoulder Pyrocarbon hemiarthroplasties have gained popularity in recent years and provide an exciting new avenue in shoulder surgery. However, we present a case report of catastrophic failure of a Pyrocarbon hemiarthroplasty due to mechanical abrasion from a residual metal anchor in the patients glenoid. We believe all surgeons using Pyrocarbon implants, at any body site, should be aware of this avoidable complication, and urge all metallic implants to be removed before insertion of a Pyrocarbon prosthesis.
Cardiac Structural Anomalies and RepairOrthopaedic implants and arthroplastyShoulder and Clavicle InjuriesMaterials scienceForensic engineeringEngineering
Citations
0
FWCI
0.00
field-weighted impact
References
0
Percentile
18%
vs. same field & year
Citation Network
How this paper connects to the literature. Drag to explore, click any node to open that paper.