Scinovex
article

Dislocations after total hip-replacement arthroplasties.

Journal of Bone and Joint Surgery · 1978 · Vol. 60(2) · pp. 217–220

Abstract

In a series of 300 total hip replacements, nine (3 per cent) dislocated. Precise measurements of the orientation of the acetabular cup were made and it was found that anterior dislocations were associated with increased acetabular-component anteversion. There was no significant correlation between cup-orientation angle and posterior dislocation. The dislocation rate for cup orientation with anteversion of 15 +/- 10 degrees and lateral opening of 40 +/- 10 degrees was 1.5 per cent, while outside this "safe" range the dislocation rate was 6.1 per cent. Other factors that were documented include time after surgery (with the greatest risk in the first thirty days) and surgical history (with a greater risk in hips that have had prior surgery).

Orthopaedic implants and arthroplastyTotal Knee Arthroplasty OutcomesHip disorders and treatmentsMedicineDislocationTotal hip replacementOrientation (vector space)SurgeryAcetabulumOrthodonticsMaterials scienceMathematicsGeometry
Citations
2,796
FWCI
1.94
field-weighted impact
References
0
Percentile
88%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.