Scinovex
article Open Access

Evaluation of stress distribution in bone at varying lengths and diameters of micro-implant

International Journal of Applied Dental Sciences · 2020 · Vol. 6(2) · pp. 68–73

Abstract

Introduction: Temporary Anchorage Devices play an important role to overcome the limitations of conventional anchorage control in fixed orthodontic practice.Aims and Objectives: To evaluate the stress distribution in bone at varying lengths and diameters of the micro-implant. Material and Methods: 3-D Finite element models of maxilla, mandible, and micro-implants with varying lengths and diameters were generated. The micro-implants were inserted 90ᵒ to the bone surface. A force of 200 g was applied from the micro-implant to the power arm. Stress distribution and its magnitude were then analyzed with a 3-dimensional finite element analysis program. Results: Increased diameter of micro-implant, showed significant negative correlation with the stress generated. The maximum von Mises stress was found for implant of diameter 1.2 mm and least for implant of diameter 1.8 mm. Conclusion: Increasing the diameter of the micro-implant reduces the stress concentration in bone, thereby increasing the likelihood of implant stabilization.

Dental Implant Techniques and OutcomesOrthodontics and Dentofacial OrthopedicsDental Radiography and Imagingvon Mises yield criterionImplantMaterials scienceFinite element methodStress (linguistics)Mandible (arthropod mouthpart)DentistryBiomedical engineeringOrthodonticsStructural engineering
Citations
0
FWCI
0.00
field-weighted impact
References
0
Percentile
29%
vs. same field & year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Evaluation of stress distribution in bone at varying lengths and diameters of micro-implant · Scinovex