Scinovex
reviewTop 1% cited

Receiver operating characteristic (ROC) curves: review of methods with applications in diagnostic medicine

Physics in Medicine and Biology · 2018 · Vol. 63(7) · pp. 07TR01–07TR01
Nancy A. ObuchowskiJennifer Bullen

Abstract

Receiver operating characteristic (ROC) analysis is a tool used to describe the discrimination accuracy of a diagnostic test or prediction model. While sensitivity and specificity are the basic metrics of accuracy, they have many limitations when characterizing test accuracy, particularly when comparing the accuracies of competing tests. In this article we review the basic study design features of ROC studies, illustrate sample size calculations, present statistical methods for measuring and comparing accuracy, and highlight commonly used ROC software. We include descriptions of multi-reader ROC study design and analysis, address frequently seen problems of verification and location bias, discuss clustered data, and provide strategies for testing endpoints in ROC studies. The methods are illustrated with a study of transmission ultrasound for diagnosing breast lesions.

AI in cancer detectionReliability and Agreement in MeasurementBreast Lesions and CarcinomasReceiver operating characteristicComputer scienceSensitivity (control systems)Diagnostic accuracySample size determinationSoftwareArtificial intelligencePattern recognition (psychology)StatisticsData mining

MeSH terms

Breast NeoplasmsFemaleHumansResearch DesignROC CurveSoftwareUltrasonography, MammarySample Size
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.