Scinovex
article Open AccessTop 10% cited

Non-invasive assessment of local arterial pulse pressure: comparison of applanation tonometry and echo-tracking

Journal of Hypertension · 2001 · Vol. 19(6) · pp. 1037–1044
Luc M. Van BortelElisabeth J. BalkesteinJanneke J. van der Heijden-SpekFloris VanmolkotJan A. StaessenJohannes A. KragtenJ.W. VredeveldMichel E. SafarHarry A. Struijker BoudierArnold P.G. Hoeks

Abstract

Pulse pressure obtained from alternatively calibrated tonometer-derived pressure waves and echo-tracking-derived distension waves demonstrates good accuracy. Accuracy of pulse pressure from internally calibrated applanation tonometer readings at the carotid artery is poor.

Cardiovascular Health and Disease PreventionHemodynamic Monitoring and TherapyBlood Pressure and Hypertension StudiesPulse pressureMedicineBlood pressureApplanation tonometryCalibrationPulse (music)Arterial treeDistensionPressure measurementPulse wave velocity

MeSH terms

AdultAgedArteriesBlood PressureDiagnostic Techniques, CardiovascularFemaleHumansMaleMiddle AgedPulsatile FlowUltrasonography