Evaluation of Multidimensional Chromatography Coupled with Tandem Mass Spectrometry (LC/LC−MS/MS) for Large-Scale Protein Analysis:  The Yeast Proteome

Journal of Proteome Research · 2002 · Vol. 2(1) · pp. 43–50
Junmin PengJoshua E. EliasCarson C. ThoreenLarry LickliderSteven P. Gygi

Abstract

Highly complex protein mixtures can be directly analyzed after proteolysis by liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS). In this paper, we have utilized the combination of strong cation exchange (SCX) and reversed-phase (RP) chromatography to achieve two-dimensional separation prior to MS/MS. One milligram of whole yeast protein was proteolyzed and separated by SCX chromatography (2.1 mm i.d.) with fraction collection every minute during an 80-min elution. Eighty fractions were reduced in volume and then re-injected via an autosampler in an automated fashion using a vented-column (100 microm i.d.) approach for RP-LC-MS/MS analysis. More than 162,000 MS/MS spectra were collected with 26,815 matched to yeast peptides (7,537 unique peptides). A total of 1,504 yeast proteins were unambiguously identified in this single analysis. We present a comparison of this experiment with a previously published yeast proteome analysis by Yates and colleagues (Washburn, M. P.; Wolters, D.; Yates, J. R., III. Nat. Biotechnol. 2001, 19, 242-7). In addition, we report an in-depth analysis of the false-positive rates associated with peptide identification using the Sequest algorithm and a reversed yeast protein database. New criteria are proposed to decrease false-positives to less than 1% and to greatly reduce the need for manual interpretation while permitting more proteins to be identified.

Advanced Proteomics Techniques and ApplicationsMass Spectrometry Techniques and ApplicationsMetabolomics and Mass Spectrometry StudiesChromatographyChemistryTandem mass spectrometryProteomeMass spectrometryYeastElutionLiquid chromatography–mass spectrometryProteomicsPeptide

MeSH terms

CationsChromatography, High Pressure LiquidChromatography, Ion ExchangeChromatography, LiquidFalse Positive ReactionsFungal ProteinsIonsPeptidesSaccharomyces cerevisiaeMass SpectrometryTime FactorsDatabases as TopicProteomeNanotechnology
Citations
1,677
FWCI
43.49
field-weighted impact
References
25
Percentile
100%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.