Scinovex
article Open AccessTop 1% cited

The management of respiratory motion in radiation oncology report of AAPM Task Group 76a)

Medical Physics · 2006 · Vol. 33(10) · pp. 3874–3900
Paul KeallG MagerasJames M. BalterRichard EmeryKenneth ForsterSteve JiangJ M KapatoesDaniel A. LowMartin J. MurphyBrad MurrayChester R. RamseyMarcel van HerkSastry VedamJohn WongEllen Yorke

Abstract

This document is the report of a task group of the AAPM and has been prepared primarily to advise medical physicists involved in the external-beam radiation therapy of patients with thoracic, abdominal, and pelvic tumors affected by respiratory motion. This report describes the magnitude of respiratory motion, discusses radiotherapy specific problems caused by respiratory motion, explains techniques that explicitly manage respiratory motion during radiotherapy and gives recommendations in the application of these techniques for patient care, including quality assurance (QA) guidelines for these devices and their use with conformal and intensity modulated radiotherapy. The technologies covered by this report are motion-encompassing methods, respiratory gated techniques, breath-hold techniques, forced shallow-breathing methods, and respiration-synchronized techniques. The main outcome of this report is a clinical process guide for managing respiratory motion. Included in this guide is the recommendation that tumor motion should be measured (when possible) for each patient for whom respiratory motion is a concern. If target motion is greater than 5 mm, a method of respiratory motion management is available, and if the patient can tolerate the procedure, respiratory motion management technology is appropriate. Respiratory motion management is also appropriate when the procedure will increase normal tissue sparing. Respiratory motion management involves further resources, education and the development of and adherence to QA procedures.

Advanced Radiotherapy TechniquesRadiation Therapy and DosimetryLung Cancer Diagnosis and TreatmentRadiation therapyMedicineQuality assuranceRespiratory systemMedical physicsRespiratory monitoringMotion (physics)RadiologyComputer scienceInternal medicine

MeSH terms

HumansMotionNeoplasmsQuality ControlRadiometryRadiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiotherapy, Computer-AssistedRespirationTime FactorsTomography, X-Ray ComputedGuidelines as TopicRadiation OncologyRadiotherapy, Intensity-Modulated
Citations
2,235
FWCI
82.33
field-weighted impact
References
211
Percentile
100%
vs. same field & year
Citations per year
References
Radiation pneumonitis as a function of mean lung dose: an analysis of pooled data of 540 patients
International Journal of Radiation Oncology*Biology*Physics · 1998 · 751 citations
Motion adaptive x-ray therapy: a feasibility study
Physics in Medicine and Biology · 2000 · 524 citations
Organ motion and its management
International Journal of Radiation Oncology*Biology*Physics · 2001 · 940 citations
Physical aspects of a real-time tumor-tracking system for gated radiotherapy
International Journal of Radiation Oncology*Biology*Physics · 2000 · 658 citations
The use of active breathing control (ABC) to reduce margin for breathing motion
International Journal of Radiation Oncology*Biology*Physics · 1999 · 911 citations
Citation Network