Recommendations for Improving Serum Creatinine Measurement: A Report from the Laboratory Working Group of the National Kidney Disease Education Program

Clinical Chemistry · 2005 · Vol. 52(1) · pp. 5–18
Gary L. Myers

Abstract

The current variability in serum creatinine measurements renders all estimating equations for GFR, including the MDRD Study equation, less accurate in the normal and slightly increased range of serum creatinine concentrations [<133 micromol/L (1.5 mg/dL)], which is the relevant range for detecting CKD [<60 mL.min(-1).(1.73 m2)(-1)]. Many automated routine methods for serum creatinine measurement meet or exceed the required precision; therefore, reduction of analytical bias in creatinine assays is needed. Standardization of calibration does not correct for analytical interferences (nonspecificity bias). The bias and nonspecificity problems associated with some of the routine methods must be addressed.

Chronic Kidney Disease and DiabetesAcute Kidney Injury ResearchDialysis and Renal Disease ManagementCreatinineRenal functionKidney diseaseMedicineStandardizationUrologyIntensive care medicineInternal medicineComputer science

MeSH terms

CalibrationChronic DiseaseCreatinineGlomerular Filtration RateHumansKidney DiseasesMathematicsReference StandardsReference ValuesPractice Guidelines as TopicClinical Chemistry TestsSerum