A frameless stereotaxic integration of computerized tomographic imaging and the operating microscope

Journal of neurosurgery · 1986 · Vol. 65(4) · pp. 545–549

Abstract

A computer-based system has been developed for the integration and display of computerized tomography (CT) image data in the operating microscope in the correct perspective without requiring a stereotaxic frame. Spatial registration of the CT image data is accomplished by determination of the position of the operating microscope as its focal point is brought to each of three CT-imaged fiducial markers on the scalp. Monitoring of subsequent microscope positions allows appropriate reformatting of CT data into a common coordinate system. The position of the freely moveable microscope is determined by a non-imaging ultrasonic range-finder consisting of three spark gaps attached to the microscope and three microphones on a rigid support in the operating room. Measurement of the acoustic impulse transit times from the spark gaps to the microphones enables calculation of those distances and unique determination of the microscope position. The CT data are reformatted into a plane and orientation corresponding to the microscope's focal plane or to a deeper parallel plane if required. This reformatted information is then projected into the optics of the operating microscope using a miniature cathode ray tube and a beam splitter. The operating surgeon sees the CT information (such as a tumor boundary) superimposed upon the operating field in proper position, orientation, and scale.

Surgical Simulation and TrainingAdvanced Radiotherapy TechniquesAugmented Reality ApplicationsMicroscopeOperating microscopeTomographyOpticsOrientation (vector space)Fiducial markerMicroscopyArtificial intelligenceComputer sciencePhysics

MeSH terms

Evaluation Studies as TopicFemaleHumansMicroscopyMiddle AgedStereotaxic TechniquesSurgical EquipmentTomography, X-Ray Computed
Citations
656
FWCI
3.16
field-weighted impact
References
4
Percentile
92%
vs. same field & year
Citations per year
Cited by
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

A frameless stereotaxic integration of computerized tomographic imaging and the operating microscope · Scinovex