Scinovex
articleTop 10% cited

Attenuation correction for a combined 3D PET/CT scanner

Medical Physics · 1998 · Vol. 25(10) · pp. 2046–2053
Paul E. KinahanD.W. TownsendThomas BeyerD. Sashin

Abstract

In this work we demonstrate the proof of principle of CT-based attenuation correction of 3D positron emission tomography (PET) data by using scans of bone and soft tissue equivalent phantoms and scans of humans. This method of attenuation correction is intended for use in a single scanner that combines volume-imaging (3D) PET with x-ray computed tomography (CT) for the purpose of providing accurately registered anatomical localization of structures seen in the PET image. The goal of this work is to determine if we can perform attenuation correction of the PET emission data using accurately aligned CT attenuation information. We discuss possible methods of calculating the PET attenuation map at 511 keV based on CT transmission information acquired from 40 keV through 140 keV. Data were acquired on separate CT and PET scanners and were aligned using standard image registration procedures. Results are presented on three of the attenuation calculation methods: segmentation, scaling, and our proposed hybrid segmentation/scaling method. The results are compared with those using the standard 3D PET attenuation correction method as a gold standard. We demonstrate the efficacy of our proposed hybrid method for converting the CT attenuation map from an effective CT photon energy of 70 keV to the PET photon energy of 511 keV. We conclude that using CT information is a feasible way to obtain attenuation correction factors for 3D PET.

Medical Imaging Techniques and ApplicationsAdvanced X-ray and CT ImagingRadiation Dose and ImagingCorrection for attenuationAttenuationScannerPositron emission tomographyPET-CTNuclear medicineIterative reconstructionCardiac PETTomographyPhysics

MeSH terms

BiophysicsHumansImage Processing, Computer-AssistedLiver NeoplasmsRadiographic Image EnhancementTomography, Emission-ComputedTomography, X-Ray ComputedPhantoms, ImagingBiophysical Phenomena
Citations
929
FWCI
7.46
field-weighted impact
References
21
Percentile
98%
vs. same field & year
Citations per year
Cited by
Medical Image Synthesis with Deep Convolutional Adversarial Networks
IEEE Transactions on Biomedical Engineering · 2018 · 606 citations
References
Energy-selective reconstructions in X-ray computerised tomography
Physics in Medicine and Biology · 1976 · 1,967 citations
Analytic 3D image reconstruction using all detected events
IEEE Transactions on Nuclear Science · 1989 · 761 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.