Scinovex
articleTop 1% cited

The Maia Spectroscopy Detector System: Engineering for Integrated Pulse Capture, Low-Latency Scanning and Real-Time Processing

AIP conference proceedings · 2010 · pp. 240–243
R. KirkhamPaul DunnA. KuczewskiD. P. SiddonsR. DodanwelaG. F. MoorheadC.G. RyanG. De GeronimoR. BeuttenmullerD. PinelliMarkus PfefferP. DaveyM. JensenDavid PatersonMartin D. de JongeDaryl L. HowardMichael KuselJonathan McKinlayR. GarrettI. GentleK. NugentS. Wilkins

Abstract

The Maia detector system is engineered for energy dispersive x‐ray fluorescence spectroscopy and elemental imaging at photon rates exceeding 107/s, integrated scanning of samples for pixel transit times as small as 50μs and high definition images of 108 pixels and real‐time processing of detected events for spectral deconvolution and online display of pure elemental images. The system developed by CSIRO and BNL combines a planar silicon 384 detector array, application‐specific integrated circuits for pulse shaping and peak detection and sampling and optical data transmission to an FPGA‐based pipelined, parallel processor. This paper describes the system and the underpinning engineering solutions.

Radiation Detection and Scintillator TechnologiesNuclear Physics and ApplicationsAdvanced X-ray and CT ImagingDetectorDeconvolutionPixelIntegrated circuitSpectroscopySpectrometerOpticsPlanarField-programmable gate arrayPhoton counting

Funding

  • Commonwealth Scientific and Industrial Research Organisation
Citations
185
FWCI
12.79
field-weighted impact
References
5
Percentile
99%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

The Maia Spectroscopy Detector System: Engineering for Integrated Pulse Capture, Low-Latency Scanning and Real-Time Processing · Scinovex