Scinovex
articleTop 10% cited

Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial

Journal of Bone and Joint Surgery - British Volume · 1999 · Vol. 81(5) · pp. 907–914
J BobynG. J. StackpoolS. Adam HackingMichael TänzerJan J. Krygier

Abstract

We have studied the characteristics of bone ingrowth of a new porous tantalum biomaterial in a simple transcortical canine model using cylindrical implants 5 x 10 mm in size. The material was 75% to 80% porous by volume and had a repeating arrangement of slender interconnecting struts which formed a regular array of dodecahedron-shaped pores. We performed histological studies on two types of material, one with a smaller pore size averaging 430 microm at 4, 16 and 52 weeks and the other with a larger pore size averaging 650 microm at 2, 3, 4, 16 and 52 weeks. Mechanical push-out tests at 4 and 16 weeks were used to assess the shear strength of the bone-implant interface on implants of the smaller pore size. The extent of filling of the pores of the tantalum material with new bone increased from 13% at two weeks to between 42% and 53% at four weeks. By 16 and 52 weeks the average extent of bone ingrowth ranged from 63% to 80%. The tissue response to the small and large pore sizes was similar, with regions of contact between bone and implant increasing with time and with evidence of Haversian remodelling within the pores at later periods. Mechanical tests at four weeks indicated a minimum shear fixation strength of 18.5 MPa, substantially higher than has been obtained with other porous materials with less volumetric porosity. This porous tantalum biomaterial has desirable characteristics for bone ingrowth; further studies are warranted to ascertain its potential for clinical reconstructive orthopaedics.

Bone Tissue Engineering MaterialsOrthopaedic implants and arthroplastyDental Implant Techniques and OutcomesBiomaterialMaterials scienceTantalumPorosityImplantBiomedical engineeringComposite materialFixation (population genetics)OsseointegrationChemistry

MeSH terms

AnimalsBiocompatible MaterialsDogsFemurHaversian SystemMaterials TestingMicroscopy, Electron, ScanningSurface PropertiesTantalumTensile StrengthOsseointegrationWeight-BearingProstheses and Implants
Citations
878
FWCI
4.64
field-weighted impact
References
31
Percentile
95%
vs. same field & year
Citations per year
References
Tissue response to porous tantalum acetabular cups
The Journal of Arthroplasty · 1999 · 280 citations
The Optimum Pore Size for the Fixation of Porous-Surfaced Metal Implants by the Ingrowth of Bone
Clinical Orthopaedics and Related Research · 1980 · 831 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.