Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing.

Hypertension · 1989 · Vol. 13(6_pt_1) · pp. 647–655
Gianfranco ParatiRiccardo CasadeiAntonella GroppelliMarco Di RienzoGiuseppe Mancia

Abstract

The accuracy of blood pressure values obtained by continuous noninvasive finger blood pressure recording via the FINAPRES device was evaluated by comparison with simultaneous intraarterial monitoring both at rest and during performance of tests known to induce fast and often marked changes in blood pressure. The comparison was performed in 24 normotensive or essential hypertensive subjects. The average discrepancy between finger and intra-arterial blood pressure recorded over a 30-minute rest period was 6.5 +/- 2.6 mm Hg and 5.4 +/- 2.9 mm Hg for systolic and diastolic blood pressure, respectively; a close between-method correspondence was also demonstrated by linear regression analysis. The beat-to-beat changes in finger systolic and diastolic blood pressure were on average similar to those measured intra-arterially during tests that induced a pressor or depressor response (hand-grip, cold pressor test, diving test, Valsalva maneuver, intravenous injections of phenylephrine and trinitroglycerine) as well as during tests that caused vasomotor changes without major variations in blood pressure (application of lower body negative pressure, passive leg raising). The average between-method discrepancy in the evaluation of blood pressure changes was never greater than 4.3 and 2.0 mm Hg for systolic and diastolic blood pressure, respectively; the corresponding standard deviations ranged between 4.6 and 1.6 mm Hg. Beat-to-beat computer analysis of blood pressure variability over the 30-minute rest period provided standard deviations almost identical when calculated by separate consideration of intra-arterial and finger blood pressure tracings (3.7 and 3.8 mm Hg, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)

Heart Rate Variability and Autonomic ControlBlood Pressure and Hypertension StudiesNon-Invasive Vital Sign MonitoringBlood pressureMedicineDiastoleValsalva maneuverCold pressor testBeat (acoustics)CardiologyPhotoplethysmogramHemodynamicsVasomotor

MeSH terms

AdultBlood PressureCatheters, IndwellingFemaleFingersNitroglycerinHeart RateHumansMaleMiddle AgedMonitoring, PhysiologicPlethysmographyPressoreceptorsRest

Funding

  • Academisch Medisch Centrum
Citations
902
FWCI
13.81
field-weighted impact
References
35
Percentile
99%
vs. same field & year
Citations per year
Cited by
References
Hemodynamic fluctuations and baroreflex sensitivity in humans: a beat-to-beat model
American Journal of Physiology-Heart and Circulatory Physiology · 1987 · 855 citations
Applied Regression Analysis
Technometrics · 2005 · 18,043 citations
Reflex Regulation of Arterial Pressure during Sleep in Man
Circulation Research · 1969 · 1,283 citations
Hemodynamic regulation: investigation by spectral analysis
American Journal of Physiology-Heart and Circulatory Physiology · 1985 · 1,298 citations
Applied Regression Analysis
Technometrics · 1982 · 5,826 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.