Scinovex
articleTop 10% cited

Modification of Cortical Stimulation for Motor Evoked Potentials under General Anesthesia

Neurosurgery · 1993 · Vol. 32(2) · pp. 219–226
Makoto TaniguchiC. CedzichMakoto TaniguchiC. CedzichJohannes Schramm

Abstract

This pilot study presents a possible modification of direct cortical electrical stimulation technique for the recording of motor evoked potentials under general anesthesia. The exposed primary motor cortex was stimulated by a short train of anodal rectangular pulses at high frequency (300-500 Hz), while the compound muscle action potentials were recorded from the forearm and hand muscles. When compared with the traditional way of eliciting movement of the extremities by applying a train of pulses at lower frequency (50-60 Hz), muscle responses were obtainable at an intensity of much lower charge. It is suggested that this stimulation achieves a repetitive activation of the corticomotoneuronal tract. Responses could be continuously recorded throughout surgery and seemed to respond to surgical manipulation affecting the motor pathways. This technique seems to be applicable for intraoperative monitoring of motor pathways but requires further optimization of stimulation and recording parameters before wider clinical applications are possible.

Intraoperative Neuromonitoring and Anesthetic EffectsHistory of Medical PracticeCardiac, Anesthesia and Surgical OutcomesMedicineStimulationMotor cortexEvoked potentialForearmNeurosciencePrimary motor cortexCortex (anatomy)AnesthesiaAnatomy

MeSH terms

Anesthesia, GeneralBrain MappingBrain NeoplasmsIntracranial Arteriovenous MalformationsElectric StimulationEvoked PotentialsGliomaHumansMeningeal NeoplasmsMeningiomaMotor CortexMotor NeuronsMusclesSomatosensory CortexMonitoring, Intraoperative

Funding

  • Deutsche Forschungsgemeinschaft
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Modification of Cortical Stimulation for Motor Evoked Potentials under General Anesthesia · Scinovex