Cone‐beam computed tomography with a flat‐panel imager: Initial performance characterization

Medical Physics · 2000 · Vol. 27(6) · pp. 1311–1323
David A. JaffrayJeffrey H. Siewerdsen

Abstract

The development and performance of a system for x-ray cone-beam computed tomography (CBCT) using an indirect-detection flat-panel imager (FPI) is presented. Developed as a bench-top prototype for initial investigation of FPI-based CBCT for bone and soft-tissue localization in radiotherapy, the system provides fully three-dimensional volumetric image data from projections acquired during a single rotation. The system employs a 512 x 512 active matrix of a-Si:H thin-film transistors and photodiodes in combination with a luminescent phosphor. Tomographic imaging performance is quantified in terms of response uniformity, response linearity, voxel noise, noise-power spectrum (NPS), and modulation transfer function (MTF), each in comparison to the performance measured on a conventional CT scanner. For the geometry employed and the objects considered, response is uniform to within 2% and linear within 1%. Voxel noise, at a level of approximately 20 HU, is comparable to the conventional CT scanner. NPS and MTF results highlight the frequency-dependent transfer characteristics, confirming that the CBCT system can provide high spatial resolution and does not suffer greatly from additive noise levels. For larger objects and/or low exposures, additive noise levels must be reduced to maintain high performance. Imaging studies of a low-contrast phantom and a small animal (a euthanized rat) qualitatively demonstrate excellent soft-tissue visibility and high spatial resolution. Image quality appears comparable or superior to that of the conventional scanner. These quantitative and qualitative results clearly demonstrate the potential of CBCT systems based upon flat-panel imagers. Advances in FPI technology (e.g., improved x-ray converters and enhanced electronics) are anticipated to allow high-performance FPI-based CBCT for medical imaging. General and specific requirements of kilovoltage CBCT systems are discussed, and the applicability of FPI-based CBCT systems to tomographic localization and image-guidance for radiotherapy is considered.

Medical Imaging Techniques and ApplicationsAdvanced X-ray and CT ImagingDigital Radiography and Breast ImagingScannerOptical transfer functionOpticsImaging phantomImage resolutionFlat panel detectorImage qualityCone beam computed tomographyNoise (video)Voxel

MeSH terms

AnimalsBiophysicsHumansMaleProstatic NeoplasmsRadiographic Image Interpretation, Computer-AssistedRadiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedPhantoms, ImagingRatsBiophysical Phenomena
Citations
530
FWCI
9.88
field-weighted impact
References
32
Percentile
99%
vs. same field & year
Citations per year
Cited by
Flat-panel cone-beam computed tomography for image-guided radiation therapy
International Journal of Radiation Oncology*Biology*Physics · 2002 · 1,322 citations
Respiratory correlated cone beam CT
Medical Physics · 2005 · 658 citations
Flat-detector computed tomography (FD-CT)
European Radiology · 2007 · 301 citations
References
Multi‐slice helical CT: Scan and reconstruction
Medical Physics · 1999 · 496 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.