Line-imaging velocimeter for shock diagnostics at the OMEGA laser facility
Abstract
A line-imaging velocity interferometer has been implemented at the OMEGA laser facility of the Laboratory for Laser Energetics, University of Rochester. This instrument is the primary diagnostic for a variety of experiments involving laser-driven shock-wave propagation, including high-pressure equation of state experiments, materials characterization experiments, shock characterization for Rayleigh–Taylor experiments, and shock timing experiments for inertial confinement fusion research. Using a laser probe beam to illuminate a target, the instrument measures shock breakout times at temporal resolutions as low as 20 ps, and spatial resolution ∼4 μm. For velocity measurements the detection limit is <0.1 km/s, and velocities of interfaces, free surfaces, and shock fronts traveling through transparent media can be measured with accuracies ∼1% over the range from 4 km/s to greater than 50 km/s. Quantitative measurements of the optical reflectance of ionizing shock fronts can also be obtained simultaneously with the velocity measurements.
How this paper connects to the literature. Drag to explore, click any node to open that paper.