Abstract

Heart rate monitoring using wrist-type photoplethysmographic signals during subjects' intensive exercise is a difficult problem, since the signals are contaminated by extremely strong motion artifacts caused by subjects' hand movements. So far few works have studied this problem. In this study, a general framework, termed TROIKA, is proposed, which consists of signal decomposiTion for denoising, sparse signal RecOnstructIon for high-resolution spectrum estimation, and spectral peaK trAcking with verification. The TROIKA framework has high estimation accuracy and is robust to strong motion artifacts. Many variants can be straightforwardly derived from this framework. Experimental results on datasets recorded from 12 subjects during fast running at the peak speed of 15 km/h showed that the average absolute error of heart rate estimation was 2.34 beat per minute, and the Pearson correlation between the estimates and the ground truth of heart rate was 0.992. This framework is of great values to wearable devices such as smartwatches which use PPG signals to monitor heart rate for fitness.

Non-Invasive Vital Sign MonitoringHeart Rate Variability and Autonomic ControlECG Monitoring and AnalysisPhotoplethysmogramHeart rateSmartwatchWristWearable computerComputer scienceSIGNAL (programming language)Ground truthArtificial intelligenceSpeech recognition

MeSH terms

HumansMalePattern Recognition, AutomatedPhysical EnduranceRunningSensitivity and SpecificityWristReproducibility of ResultsArtifactsPhotoplethysmographyMonitoring, AmbulatoryYoung AdultAdolescentAdultAlgorithms
Citations
729
FWCI
29.17
field-weighted impact
References
36
Percentile
100%
vs. same field & year
Citations per year
References
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
IEEE Transactions on Instrumentation and Measurement · 2011 · 335 citations
Sparse signal reconstruction from limited data using FOCUSS: a re-weighted minimum norm algorithm
IEEE Transactions on Signal Processing · 1997 · 1,821 citations
Regression Shrinkage and Selection Via the Lasso
Journal of the Royal Statistical Society Series B (Statistical Methodology) · 1996 · 50,746 citations
Sparse solutions to linear inverse problems with multiple measurement vectors
IEEE Transactions on Signal Processing · 2005 · 1,402 citations
Compressed sensing
IEEE Transactions on Information Theory · 2006 · 22,859 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

TROIKA: A General Framework for Heart Rate Monitoring Using Wrist-Type Photoplethysmographic Signals During Intensive Physical Exercise · Scinovex