Within-Day Reproducibility of an HPLC−MS-Based Method for Metabonomic Analysis:  Application to Human Urine

Journal of Proteome Research · 2007 · Vol. 6(8) · pp. 3291–3303
Helen GikaGeorgios TheodoridisJulia E. WingateIan D. Wilson

Abstract

Self-evidently, research in areas supporting "systems biology" such as genomics, proteomics, and metabonomics are critically dependent on the generation of sound analytical data. Metabolic phenotyping using LC-MS-based methods is currently at a relatively early stage of development, and approaches to ensure data quality are still developing. As part of studies on the application of LC-MS in metabonomics, the within-day reproducibility of LC-MS, with both positive and negative electrospray ionization (ESI), has been investigated using a standard "quality control" (QC) sample. The results showed that the first few injections on the system were not representative, and should be discarded, and that reproducibility was critically dependent on signal intensity. On the basis of these findings, an analytical protocol for the metabonomic analysis of human urine has been developed with proposed acceptance criteria based on a step-by-step assessment of the data. Short-term sample stability for human urine was also assessed. Samples were stable for at least 20 h at 4 degrees C in the autosampler while queuing for analysis. Samples stored at either -20 or -80 degrees C for up to 1 month were indistinguishable on subsequent LC-MS analysis. Overall, by careful monitoring of the QC data, it is possible to demonstrate that the "within-day" reproducibility of LC-MS is sufficient to ensure data quality in global metabolic profiling applications.

Metabolomics and Mass Spectrometry StudiesAdvanced Chemical Sensor TechnologiesMass Spectrometry Techniques and ApplicationsReproducibilityUrineChromatographySample preparationProtocol (science)Electrospray ionizationMetabolomicsComputer scienceChemistryMass spectrometry

MeSH terms

Chromatography, High Pressure LiquidHumansUrineReproducibility of ResultsProteomeSpectrometry, Mass, Electrospray Ionization
Citations
508
FWCI
10.32
field-weighted impact
References
14
Percentile
99%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Within-Day Reproducibility of an HPLC−MS-Based Method for Metabonomic Analysis:  Application to Human Urine · Scinovex