Abstract

Over the past two decades, spinal cord stimulation devices and techniques have evolved from single-channel systems, with electrodes requiring laminectomy, into programmable "multichannel" systems with electrodes that may be placed percutaneously. We have reviewed our experience in 320 consecutive patients treated with these devices at our institution between 1972 and 1990. Technical details of treatment as well as patient characteristics have been assessed as predictors of clinical outcome and of hardware reliability by univariate and multivariate statistical methods. Current follow-up has been obtained at intervals from 2 to 20 years (mean, 7.1 yr) postoperatively on 205 patients. All clinical outcome measures have been based on disinterested third-party interview data--standard analog pain ratings, employment status, activities of daily living, and use of analgesics. At 7-year mean follow-up, 52% of the 171 patients who received permanent implants reported at least 50% continued pain relief. A majority had maintained improvements in activities of daily living and analgesic use. Analysis of hardware reliability for 298 permanent implants revealed significantly fewer clinical failures (P < 0.001) and technical failures (in particular, electrode migration and malposition, P = 0.025) as single-channel implants have evolved into programmable, multichannel devices. Our analysis of technical and clinical prognostic factors may be useful to the clinician in selecting patients for this procedure.

Pain Management and TreatmentMusculoskeletal pain and rehabilitationTranscranial Magnetic Stimulation StudiesMedicineSpinal cord stimulatorUnivariate analysisIntractable painLaminectomyMultivariate analysisSurgerySpinal cord stimulationAnalgesicActivities of daily living

MeSH terms

AdultAgedAged, 80 and overElectrodes, ImplantedElectric Stimulation TherapyEquipment DesignEquipment FailureFemaleFollow-Up StudiesHumansLaminectomyMaleMicrocomputersMiddle AgedPain Measurement
Citations
493
FWCI
9.74
field-weighted impact
References
95
Percentile
99%
vs. same field & year
Citations per year
References
Studies with pain rating scales.
Annals of the Rheumatic Diseases · 1978 · 1,951 citations
Statistical Aspects of the Analysis of Data From Retrospective Studies of Disease
JNCI Journal of the National Cancer Institute · 1959 · 14,848 citations
Regression Models and Life-Tables
Journal of the Royal Statistical Society Series B (Statistical Methodology) · 1972 · 38,926 citations
Nonparametric Estimation from Incomplete Observations
Journal of the American Statistical Association · 1958 · 38,793 citations
Electrical Inhibition of Pain by Stimulation of the Dorsal Columns
Anesthesia & Analgesia · 1967 · 994 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.