Precise radiochromic film dosimetry using a flat-bed document scanner

Medical Physics · 2005 · Vol. 32(7Part1) · pp. 2245–2253
Slobodan DevićJan SeuntjensE ShamErvin B. PodgoršakC. Ross SchmidtleinAssen S. KirovChristopher G. Soares

Abstract

In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic® film models, HS and Prototype A EBT exposed to 6MV photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around 0.4Gy in the case of the EBT (Prototype A), and for doses above 5Gy in the case of the HS model GafChromic® film using a region of interest 2×2mm2 in size.

Advanced Radiotherapy TechniquesMedical Imaging Techniques and ApplicationsRadiation Detection and Scintillator TechnologiesDosimetryScannerMedical physicsMedical imagingNuclear medicineOpticsMaterials scienceComputer sciencePhysicsMedicine

MeSH terms

AlgorithmsDose-Response Relationship, RadiationFilm DosimetryRadiation DosageComputer Peripherals

Funding

  • National Cancer Institute
Citations
568
FWCI
25.19
field-weighted impact
References
32
Percentile
100%
vs. same field & year
Citations per year
Cited by
Multichannel film dosimetry with nonuniformity correction
Medical Physics · 2011 · 448 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.