articleTop 1% cited
Spin warp NMR imaging and applications to human whole-body imaging
Physics in Medicine and Biology · 1980 · Vol. 25(4) · pp. 751–756
W. A. Edelstein✉(University of Aberdeen)J.M.S. Hutchison(University of Aberdeen)G. Allan Johnson(University of Aberdeen)T W Redpath(University of Aberdeen)
Abstract
Describes a new nuclear magnetic resonance (NMR) imaging technique which the authors call 'spin warp imaging' and gives examples of its application to human whole-body imaging. The apparatus is based on a four-coil, air cored magnet (made by the Oxford Instrument Company) capable of accepting the whole human body. The magnet produces a static field of 0.04 T giving a proton NMR frequency of 1.7 MHz. The maximum field inhomogeneity is about 6*10-4 at a radius of 0.23 m, approximately twice the amount theoretically attainable with this configuration. The pulse sequence used is shown.
Advanced MRI Techniques and ApplicationsAtomic and Subatomic Physics ResearchNMR spectroscopy and applicationsMagnetNuclear magnetic resonancePulse sequenceRADIUSSpin (aerodynamics)Electromagnetic coilPhysicsSpin echoMagnetic resonance imagingMaterials science
MeSH terms
HumansMagnetic Resonance SpectroscopyTomography