Scinovex
articleTop 1% cited

Specific volume measurements of Cu, Mo, Pd, and Ag and calibration of the ruby <i>R</i>1 fluorescence pressure gauge from 0.06 to 1 Mbar

Journal of Applied Physics · 1978 · Vol. 49(6) · pp. 3276–3283
Ho‐kwang MaoP. M. BellJ. W. ShanerDaniel Steinberg

Abstract

The wavelength shift with pressure of the ruby R1 fluorescence line (Δλ) has been calibrated in the diamond-window pressure cell from 0.06 to 1 Mbar. This was done by simultaneously making specific volume measurements of four metals (Cu, Mo, Ag, and Pd) and referring these results to isothermal equations of state derived from shock-wave experiments. The result is P (Mbar) = (19.04/5) {[(λ0+Δλ)/λ0]5−1}, where λ0 is the wavelength measured at 1 bar.

High-pressure geophysics and materialsChemical Thermodynamics and Molecular StructureSpectroscopy and Laser ApplicationsFluorescenceCalibrationBar (unit)WavelengthVolume (thermodynamics)Isothermal processPressure measurementAnalytical Chemistry (journal)Diamond anvil cellShock wave
Citations
2,240
FWCI
11.35
field-weighted impact
References
7
Percentile
99%
vs. same field & year
Citations per year
References
Related articles
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.