Scinovex
articleTop 10% cited

Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses

Journal of Bone and Joint Surgery - British Volume · 1994 · Vol. 76-B(6) · pp. 912–917

Abstract

Roentgen stereophotogrammetry was used to measure the migration of the centre of the femoral head in 84 cemented Lubinus SP I hip arthroplasties (58 primary operations, 26 revisions). Four to seven years later, seven femoral components had been revised because of painful loosening. These implants showed greater subsidence, medial migration and posterior migration during the first two postoperative years than did the hips which had not been revised. Six months after operation, subsidence of more than 0.33 mm combined with a total migration of more than 0.85 mm predicted an increased risk of subsequent revision; the amount of subsidence at two years was an even better predictor. The probability of revision was greater than 50% if the subsidence at two years was 1.2 mm or more.

Orthopaedic implants and arthroplastyTotal Knee Arthroplasty OutcomesHip disorders and treatmentsMedicineSubsidenceFemoral headTotal hip replacementProsthesisSurgeryDentistryOrthodonticsGeology

MeSH terms

PrognosisAdultAgedAged, 80 and overAnalysis of VarianceBone CementsFemaleFemur HeadFollow-Up StudiesHip ProsthesisHumansMaleMiddle AgedOsteolysisPhotogrammetry
Citations
619
FWCI
7.89
field-weighted impact
References
33
Percentile
98%
vs. same field & year
Citations per year
References
Impacted cancellous allografts and cement for revision total hip arthroplasty
Journal of Bone and Joint Surgery - British Volume · 1993 · 767 citations
Related articles
Study of clinical and functional outcome of total hip replacement in avascular necrosis of femoral head
International Journal of Orthopaedics Sciences · 2018 · 5 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.

Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses · Scinovex